This recall involves updating instructions for using devices, and does not involve removing them from where they are used or sold. The FDA has identified this recall as the most serious type. This device may cause serious injury or death if you continue to use it without following the updated instructions.  

Affected Product  

  • Product Names: Philips Allura and Azurion systems used with a Philips mattress
  • Unique Device Identifier (UDI)/Model: See full list of affected devices below. 
  • Lot/Serial Numbers: See full list of affected devices below.   
     

What to Do  

  • Follow additional instructions provided by Philips as an addendum to the instructions for use and as a Quick Reference Card.

On January 13, 2025, Philips sent all affected customers an Urgent Medical Device Correction recommending the following actions:

  • Circulate the letter to all users so they are aware of the issue.
  • Follow additional instructions on the use of the mattress, which will be provided as an addendum to the Instructions for Use and a Quick Reference Card. The instructions were attached to the letter as Appendix B and include: 
    • Instructions for the use of the Mattress
      • Transfer of patient
        • Communicate clearly with the other hospital staff during the transfer process to ensure that everyone is aware of the patient’s weight and position.
        • Open the air plug of the mattress. Opening the air plug allows the mattress to expand and contract with the weight of the patient.
        • Ensure that patient transfer devices are used correctly.
        • Move the patient carefully and avoid movements that could cause the mattress to slip. 
        • If a patient positions themselves on the table, ensure that the patient knows that the mattress is NOT fixed to the table.
  • Use of correct mattress and positioning of the mattress on the tabletop
    • Ensure that the appropriate mattress is used on the table.
  • Be aware that the mattress is placed on the tabletop with the table surface completely supporting the mattress.
  • Please be aware of the narrow area of the neuro tabletop. Do not place the mattress where the tabletop is not underneath the mattress.
  • Download the Instructions for Use addendum and Quick Reference Card in multiple languages at: https://www.usa.philips.com/healthcare/about/support/resource-center
  • Send a copy of the letter to any organization that may have received these devices and report this information to Philips by email at IGT_Recalls@philips.com.
  • Complete the response form attached to the letter within 30 days of receiving it.
  • Report any of these issues if they occur to the local Philips representative at 1-800-722-9377.

Reason for Updates to Use Instructions

Philps is updating use instructions for Allura and Azurion system patient tables after identifying some situations that may result in the patient falling from the table. These situations include: the mattress slipping from the table during patient transfer, incorrect positioning of the neuro mattress on the table, or if the incorrect mattress is being used on the table.

The use of affected product may cause serious adverse health consequences, including bruises (contusions), bleeding under the skin outside of blood vessels (hematoma), scratches, skin abrasions, stiffness, bleeding in the skull (intracranial hemorrhage), large and/or complex cuts (lacerations), and death.

There have been five reported injuries. There have been no reports of death. 

Device Use

Philips Allura and Azurion systems are types of patient tables used in operating rooms. 

Contact Information  

Customers in the U.S. with questions about this recall should contact their local Philips representative at 1-800-722-9377 (Philips Customer Care Solutions). 

Full List of Affected Devices

Commercial Name  System Code UDI
Allura CV20 722031
Allura Xper FD10 722003 00884838059191(10)
  722010 00884838059030(21)
  722026  00884838054189(21)
 AlluraXper FD10/10 722005 00884838059191(10)
00884838080805(10)
  722011
  722027 00884838054196(21)
Allura Xper FD10C 722001
AlluraXper FD20 722006  00884838055513(21)
00884838059115(21)
00884838059191(10)
  722012 00884838059054(21)
  722028 00884838054202(21)
AlluraXper FD20 Biplane 722008 00884838059191(10)
  722013 00884838059061(21)
AlluraXper FD20 Biplane OR Table 722020
  722025
AlluraXper FD20 OR Table 722023
  722035
AlluraXperFD20/10 722029 00884838054219(21)
Allura Xper FD20/20 722038  00884838054226(21)
ALluraXperFD20/15  722058 00884838059115(21)
Azurion 3 M12 722063 00884838085275(21)
Azurion 3 M15 722221 00884838099203(21)
  722064 00884838085282(21)
  722222 00884838099210(21)
  722280 00884838103276(21)
Azurion 5 M12 722227 00884838099227(21)
  722231 00884838116740(21)
Azurion 5 M20 722228  00884838099234(21)
  722232 00884838116757(21)
  722281 00884838110564(21)
Azurion 7 B12 722067 00884838085350(21)
  722225 00884838099265(21)
  722235 00884838116788(21)
Azurion 7 B20 722068 00884838085367(21)
  722226 00884838099272(21)
  722236 00884838116801(21)
Azurion 7 M12 722078 00884838085251(21)
  722223 00884838099241(21)
  722233 00884838116764(21)
Azurion 7 M20 722224  00884838099258(21)
  722079 00884838085268(21)
  722234 00884838116771(21)
  722282 00884838110571(21)
Cardiovascular-Allura Centron 722400 00884838042315(21)

Additional FDA Resources  

Unique Device Identifier (UDI)

The unique device identifier (UDI) helps identify individual medical devices sold in the United States from manufacturing through distribution to patient use. The UDI allows for more accurate reporting, reviewing, and analyzing of adverse event reports so that devices can be identified, and problems potentially corrected more quickly.

How do I report a problem?

Health care professionals and consumers may report adverse reactions or quality problems they experienced using these devices to MedWatch: The FDA Safety Information and Adverse Event Reporting Program. 

  • Content current as of:

    02/28/2025

Source